Advertisement
Oral and maxillofacial radiology| Volume 115, ISSUE 4, P558-564, April 2013

Influence of anatomical location on CT numbers in cone beam computed tomography

      Objective

      To assess the influence of anatomical location on computed tomography (CT) numbers in mid- and full field of view (FOV) cone beam computed tomography (CBCT) scans.

      Study Design

      Polypropylene tubes with varying concentrations of dipotassium hydrogen phosphate (K2HPO4) solutions (50-1200 mg/mL) were imaged within the incisor, premolar, and molar dental sockets of a human skull phantom. CBCT scans were acquired using the NewTom 3G and NewTom 5G units. The CT numbers of the K2HPO4 phantoms were measured, and the relationship between CT numbers and K2HPO4 concentration was examined. The measured CT numbers of the K2HPO4 phantoms were compared between anatomical sites.

      Results

      At all six anatomical locations, there was a strong linear relationship between CT numbers and K2HPO4 concentration (R 2 >0.93). However, the absolute CT numbers varied considerably with the anatomical location.

      Conclusion

      The relationship between CT numbers and object density is not uniform through the dental arch on CBCT scans.
      We show that the relationship between CT numbers and x-ray attenuation is not uniform throughout the dental arch on CBCT scans. Given this nonuniformity, comparison of numerical gray values measured at different anatomical locations is misleading.
      Cone beam computed tomography (CBCT) is an advanced imaging modality with several applications in dentomaxillofacial diagnosis and treatment planning.
      • Tetradis S.
      • White S.C.
      A decade of cone beam computed tomography.
      Compared with conventional two-dimensional radiography, CBCT offers several advantages including visualizing the anatomic region in all three dimensions and producing images without geometric distortion and magnification. In this technique, a cone-shaped x-ray beam and a detector rotate around the object acquiring multiple projections, which are reconstructed into a volumetric image. Currently used image receptors include image intensifier and flat panel detector. Many CBCT units allow collimation of the x-ray beam to limit the amount of tissue imaged. The imaged field of view (FOV) is typically described as small (or limited), medium, or large, depending on the anatomical coverage. Typically, as the FOV increases, the radiation dose increases. Importantly, the radiation dose from CBCT is significantly lower than that from multi-slice computed tomography (MSCT) examinations.
      • White S.C.
      • Mallya S.M.
      Update on the biological effects of ionizing radiation, relative dose factors and radiation hygiene.
      In computed tomography (CT) data, each voxel in the reconstructed CT volume is represented by a numerical value termed the CT number (sometimes referred in the literature to as “gray values”). This number reflects the degree of x-ray attenuation—the average linear attenuation coefficient of that voxel. Major factors that influence the CT number include the tissue features (atomic number and density) and homogeneity and energy of the x-ray beam. In MSCT units, the CT numbers are expressed as Hounsfield unit (HU), which expresses x-ray attenuation of a voxel relative to the attenuation of water. Ideally, it would be valuable to have this CT number be also closely representative of the true x-ray attenuation of the tissue. This would be of practical value in examining the degree of mineralization of bone for implant treatment planning and in automatic segmentation of imaged volumes for computer-aided design and computer-aided manufacturing applications. Several studies have examined the relationship between CT numbers and bone quality assessment for implant treatment planning.
      • Naitoh M.
      • Aimiya H.
      • Hirukawa A.
      • Ariji E.
      Morphometric analysis of mandibular trabecular bone using cone beam computed tomography: an in vitro study.
      • Isoda K.
      • Ayukawa Y.
      • Tsukiyama Y.
      • Sogo M.
      • Matsushita Y.
      • Koyano K.
      Relationship between the bone density estimated by cone-beam computed tomography and the primary stability of dental implants.
      • Nackaerts O.
      • Maes F.
      • Yan H.
      • Couto Souza P.
      • Pauwels R.
      • Jacobs R.
      Analysis of intensity variability in multislice and cone beam computed tomography.
      • Arisan V.
      • Karabuda Z.C.
      • Avsever H.
      • Ozdemir T.
      Conventional Multi-Slice Computed Tomography (CT) and Cone-Beam CT (CBCT) for Computer-Assisted Implant Placement. Part I: Relationship of Radiographic Gray Density and Implant Stability.
      • Valiyaparambil J.V.
      • Yamany Y.
      • Ortiz D.
      • et al.
      Bone quality evaluation: comparison of CBCT and subjective surgical assessment.
      Arisan et al.
      • Arisan V.
      • Karabuda Z.C.
      • Avsever H.
      • Ozdemir T.
      Conventional Multi-Slice Computed Tomography (CT) and Cone-Beam CT (CBCT) for Computer-Assisted Implant Placement. Part I: Relationship of Radiographic Gray Density and Implant Stability.
      showed similarities between CT numbers from MSCT and CBCT in predicting primary implant stability and subjective bone quality classification.
      Unlike MSCT units, current dental CBCT units do not use a standard scaling system. Nevertheless, studies have demonstrated that the relationship between CT numbers and x-ray attenuation is linear on CBCT scans.
      • Valiyaparambil J.V.
      • Yamany Y.
      • Ortiz D.
      • et al.
      Bone quality evaluation: comparison of CBCT and subjective surgical assessment.
      • Mah P.
      • Reeves T.E.
      • McDavid W.D.
      Deriving Hounsfield units using grey levels in cone beam computed tomography.
      These two studies imaged radiographic phantoms by both CBCT and MSCT scanners to demonstrate that there is a strong correlation between CT numbers from the two modalities. Based on these data, methods have been proposed to convert CT numbers, measured on CBCT scans to HU.
      • Mah P.
      • Reeves T.E.
      • McDavid W.D.
      Deriving Hounsfield units using grey levels in cone beam computed tomography.
      • Reeves T.
      • Mah P.
      • McDavid W.
      Deriving Hounsfield units using grey levels in cone beam CT: a clinical application.
      However, such methods make the implicit assumption that the relationship between CT numbers and x-ray attenuation is uniform through the CBCT image volume. There are several factors that contribute to the inhomogeneity of CT numbers on CBCT scans. These include beam hardening, artifacts from metallic restorations, and, importantly, scattered radiation. Of particular relevance to CBCT, the amount of scattered radiation varies with the FOV, with x-ray beam parameters and also with the anatomical location.
      • Nackaerts O.
      • Maes F.
      • Yan H.
      • Couto Souza P.
      • Pauwels R.
      • Jacobs R.
      Analysis of intensity variability in multislice and cone beam computed tomography.
      However, the magnitude of the contribution of these factors to subsequent CT number inhomogeneity is not fully understood. The aim of the present study was to systematically assess the influence of FOV and anatomical location on CT numbers measured on CBCT scans.

      Materials and Methods

      Radiographic phantoms were custom-fabricated using solutions of dipotassium hydrogen phosphate (K2HPO4), a highly water-soluble salt whose effective atomic number is very close to that of calcium hydroxyapatite (15.58 and 15.86, respectively). This makes the x-ray attenuation of both substances similar. The effective linear attenuation coefficient of K2HPO4 is within 2.2%-2.7% of that for calcium hydroxyapatite over the photon energy range of 20-100 keV.
      • Sanada S.
      • Kawahara K.
      • Yamamoto T.
      • Takashima T.
      New tissue substitutes representing cortical bone and adipose tissue in quantitative radiology.
      Aqueous solutions of K2HPO4 were prepared at concentrations of 50, 200, 400, 600, 800, 1000 and 1200 mg/mL. This concentration range was selected to represent a broad range of x-ray attenuation. When scanned in a uniform field in water and air, solutions with concentrations of 1000 and 1200 mg/mL had approximately the same CT numbers as dentin and cortical bone, respectively. The range from 50 to 800 mg/mL represents attenuation of trabecular bone of varying mineralization.
      To assess the relationship between x-ray attenuation and CT numbers and the influence of anatomical location on these CT numbers, we used a human skull phantom. This phantom contained an adult human skull with a permanent dentition and the cervical spine. A uniform layer of wax surrounded the osseous structures to simulate soft tissue absorption and scatter radiation.
      • Brand J.W.
      • Kuba R.K.
      • Braunreiter T.C.
      An improved head-and-neck phantom for radiation dosimetry.
      Six teeth were carefully extracted from the skull: the maxillary and mandibular central incisors, first premolars, and first molars and replaced with polypropylene tubes containing various concentrations of K2HPO4 (Figure 1 ). CBCT scans were performed at three FOVs (6, 9, and 12 in) using the NewTom 3G unit (QR, Verona, Italy) and at two FOVs (8×8 and 18×16 cm) using the NewTom 5G unit (QR). Additionally, a high-resolution mode scan was performed at the 8×8 cm FOV. For each scan, the skull phantom was positioned to simulate the ideal position of a patient's head during a CBCT examination for the specific anatomical region.
      Figure thumbnail gr1
      Fig. 1 A, Representative section showing placement of the K2HPO4 phantom (arrow) in a tooth socket of a human skull phantom. B, Representative axial section showing measurement of the CT number within a region of the K2HPO4 phantom.
      The gray scale (bit depth) of the CBCT images was 12 bits for the NewTom 3G unit and 14 bits for the NewTom 5G unit. Volumetric data from both units were reconstructed in the native NNT software program (QR), exported to DICOM file format and imported into the OsiriX Imaging software for Mac OS X, a public domain software.
      • Rosset A.
      • Spadola L.
      • Ratib O.
      OsiriX: an open-source software for navigating in multidimensional DICOM images.
      It is important to note that the native NNT software displays the CT numbers on a scale that is analogous to the HU scale, with the minimum density (air) at −1000 HU and water at approximately zero. This scale is maintained in the DICOM format and is read accordingly by the OsiriX software. On 0.20 mm axial slices, the mean CT numbers were measured at five heights within each tube and averaged (Figure 1). Linear regression was performed to assess the relationship between CT numbers and K2HPO4 concentration. Variations in the measured CT numbers for the same K2HPO4 concentration placed at different anatomical locations and when scanned in different FOVs were determined.

      Results

      At all anatomical locations assessed on the human skull phantom, the CT numbers were discriminated between the various K2HPO4 concentrations with a strong linear relationship (R 2 >0.93). This was observed at all FOVs with both the NewTom 3G and the NewTom 5G CBCT units and at the high-resolution mode scan of the NewTom 5G unit (Figures 2 and 3 ).
      Figure thumbnail gr2
      Fig. 2Relationship between K2HPO4 concentration (mg/mL) and CT numbers. The K2HPO4 phantoms were imaged at the indicated anatomical locations using a NewTom 3G CBCT unit in 6-, 9-, or 12-in FOV. The linear regression line and the coefficient are indicated.
      Figure thumbnail gr3
      Fig. 3Relationship between K2HPO4 concentration (mg/mL) and CT numbers. The K2HPO4 phantoms were imaged at the indicated anatomical locations using a NewTom 5G CBCT unit, using the following FOVs: 18×16 cm, 8×8 cm, and 8×8 cm, high-resolution mode scan. The linear regression line and the coefficient are indicated.
      The absolute CT number varied considerably depending on the anatomical location for both the NewTom 3G and the NewTom 5G units (Figures 4 and 5 and Tables I and II ). The same concentration of K2HPO4 yielded different CT numbers when placed at different anatomical locations within the skull phantom. In general, for any given K2HPO4 concentration, the CT numbers were higher in the incisor region compared with the premolar and molar regions. This trend was observed for both the maxilla and the mandible. Furthermore, the variation in the CT numbers between the anatomical locations was observed with all FOVs examined and with both the standard and high-resolution scanning modes of the NewTom 5G unit.
      Figure thumbnail gr4
      Fig. 4Effect of anatomical location on CT numbers in the NewTom 3G CBCT unit.
      Figure thumbnail gr5
      Fig. 5Effect of anatomical location on CT numbers in the NewTom 5G CBCT unit.
      Table ICT numbers of K2HPO4 phantoms measured at the various anatomical locations of a NewTom 3G CBCT scan
      K2HPO4 (mg/mL)MaxillaMandible
      IncisorPremolarMolarIncisorPremolarMolar
      6-in FOV
       50311 ± 53148 ± 8147 ± 64347 ± 30165 ± 49120 ± 27
       200458 ± 46253 ± 73189 ± 74511 ± 27275 ± 62265 ± 31
       400580 ± 44356 ± 73353 ± 68599 ± 25470 ± 58434 ± 30
       600706 ± 47573 ± 80375 ± 55761 ± 32549 ± 58486 ± 27
       800863 ± 50699 ± 110550 ± 52931 ± 31717 ± 67637 ± 38
       10001024 ± 59829 ± 76570 ± 601062 ± 24858 ± 59845 ± 27
       12001127 ± 781005 ± 102908 ± 471292 ± 21931 ± 561053 ± 28
      9-in FOV
       50256 ± 5824 ± 80−170 ± 53379 ± 35118 ± 59−82 ± 52
       200445 ± 55156 ± 9060 ± 66505 ± 22231 ± 63144 ± 45
       400662 ± 48262 ± 89213 ± 62752 ± 32475 ± 55314 ± 46
       600799 ± 56541 ± 116231 ± 68985 ± 31590 ± 57528 ± 35
       800978 ± 55673 ± 70532 ± 521110 ± 37782 ± 45627 ± 42
       10001173 ± 77879 ± 96625 ± 651388 ± 251001 ± 64895 ± 53
       12001170 ± 861040 ± 90821 ± 501508 ± 351090 ± 57956 ± 30
      12-in FOV
       50546 ± 7044 ± 89−166 ± 79748 ± 70299 ± 82−64 ± 53
       200794 ± 68183 ± 87−27 ± 82993 ± 62425 ± 80205 ± 57
       400987 ± 86396 ± 120149 ± 781175 ± 68749 ± 55458 ± 46
       6001216 ± 79717 ± 115326 ± 801512 ± 57974 ± 54545 ± 46
       8001391 ± 66815 ± 97533 ± 641566 ± 471156 ± 63794 ± 53
       10001665 ± 681082 ± 95695 ± 791909 ± 411489 ± 671035 ± 46
       12001614 ± 731298 ± 901001 ± 682214 ± 461661 ± 571292 ± 34
      Table IICT numbers of K2HPO4 measured at the various anatomical locations of a NewTom 5G CBCT scan
      K2HPO4 (mg/mL)MaxillaMandible
      IncisorPremolarMolarIncisorPremolarMolar
      8 × 8 cm FOV
       50406 ± 4724 ± 145154 ± 95278 ± 24202 ± 36105 ± 87
       200604 ± 35388 ± 72340 ± 59538 ± 27423 ± 37378 ± 70
       400879 ± 41639 ± 82515 ± 92829 ± 38648 ± 28659 ± 49
       6001148 ± 58704 ± 122816 ± 881067 ± 35900 ± 34849 ± 71
       8001350 ± 53894 ± 1101046 ± 971227 ± 371169 ± 28974 ± 62
       10001589 ± 521114 ± 1331220 ± 1051503 ± 401438 ± 421330 ± 52
       12001754 ± 451349 ± 1071504 ± 781628 ± 271596 ± 401438 ± 36
      8 × 8 cm FOV, high-resolution mode scan
       50378 ± 6847 ± 100145 ± 74296 ± 51118 ± 53229 ± 50
       200608 ± 56228 ± 100405 ± 82524 ± 57397 ± 46399 ± 96
       400858 ± 51497 ± 99592 ± 58787 ± 59688 ± 50622 ± 72
       6001110 ± 66679 ± 119900 ± 701027 ± 41900 ± 49888 ± 58
       8001275 ± 55908 ± 891054 ± 611268 ± 651104 ± 471073 ± 51
       10001437 ± 691023 ± 901198 ± 951370 ± 631255 ± 451208 ± 64
       12001583 ± 481236 ± 751393 ± 871529 ± 711404 ± 521342 ± 48
      18 × 16 cm FOV
       50389 ± 49−18 ± 103154 ± 77357 ± 37187 ± 34213 ± 30
       200577 ± 35259 ± 70335 ± 55596 ± 45361 ± 106225 ± 132
       400835 ± 31366 ± 123518 ± 87589 ± 29694 ± 39457 ± 80
       6001044 ± 41560 ± 133724 ± 741072 ± 31925 ± 49665 ± 59
       8001258 ± 38841 ± 106916 ± 621271 ± 581118 ± 37821 ± 58
       10001382 ± 43882 ± 951263 ± 571418 ± 411258 ± 37837 ± 41
       12001537 ± 301059 ± 1101176 ± 521530 ± 491380 ± 361081 ± 23

      Discussion

      CBCT imaging has been increasingly used in dentomaxillofacial diagnosis. In addition to its ability to display images in three dimensions, several studies have explored the utility of CBCT to quantitatively assess the bone quality.
      • Naitoh M.
      • Aimiya H.
      • Hirukawa A.
      • Ariji E.
      Morphometric analysis of mandibular trabecular bone using cone beam computed tomography: an in vitro study.
      • Isoda K.
      • Ayukawa Y.
      • Tsukiyama Y.
      • Sogo M.
      • Matsushita Y.
      • Koyano K.
      Relationship between the bone density estimated by cone-beam computed tomography and the primary stability of dental implants.
      • Nackaerts O.
      • Maes F.
      • Yan H.
      • Couto Souza P.
      • Pauwels R.
      • Jacobs R.
      Analysis of intensity variability in multislice and cone beam computed tomography.
      • Arisan V.
      • Karabuda Z.C.
      • Avsever H.
      • Ozdemir T.
      Conventional Multi-Slice Computed Tomography (CT) and Cone-Beam CT (CBCT) for Computer-Assisted Implant Placement. Part I: Relationship of Radiographic Gray Density and Implant Stability.
      • Valiyaparambil J.V.
      • Yamany Y.
      • Ortiz D.
      • et al.
      Bone quality evaluation: comparison of CBCT and subjective surgical assessment.
      The practical application of this assessment is limited by the fact that different CBCT units may vary considerably in their exposure parameters and that there is no standard scaling system during image reconstruction.
      Using a K2HPO4-based phantom, we showed that the CT numbers from CBCT scans are linearly related to x-ray attenuation. This is in concordance with the previous reports.
      • Valiyaparambil J.V.
      • Yamany Y.
      • Ortiz D.
      • et al.
      Bone quality evaluation: comparison of CBCT and subjective surgical assessment.
      • Mah P.
      • Reeves T.E.
      • McDavid W.D.
      Deriving Hounsfield units using grey levels in cone beam computed tomography.
      • Nomura Y.
      • Watanabe H.
      • Honda E.
      • Kurabayashi T.
      Reliability of voxel values from cone-beam computed tomography for dental use in evaluating bone mineral density.
      • Hohlweg-Majert B.
      • Pautke C.
      • Deppe H.
      • Metzger M.C.
      • Wagner K.
      • Schulze D.
      Qualitative and quantitative evaluation of bony structures based on DICOM dataset.
      Additionally, in a pilot study performed prior to this, we observed similar linear relationship when the K2HPO4-based phantom was scanned in a uniform fields, either in water or in air. Solutions with concentrations of 1000 and 1200 mg/mL had approximately the same CT numbers as well-mineralized trabecular bone and dentin and cortical bone, respectively, of a dentate mandible scanned under the same conditions.
      Using radiographic phantoms imaged in CBCT and MSCT units, it has been shown that CT numbers also correlate well with HU. Some investigators have proposed that HU can be derived from CT numbers measured on CBCT scans, using factors that are specific for a given CBCT unit.
      • Mah P.
      • Reeves T.E.
      • McDavid W.D.
      Deriving Hounsfield units using grey levels in cone beam computed tomography.
      • Lagravère M.O.
      • Fang Y.
      • Carey J.
      • Toogood R.W.
      • Packota G.V.
      • Major P.W.
      Density conversion factor determined using a cone-beam computed tomography unit NewTom QR-DVT 9000.
      Such conversions are based on the assumption that the relationship between CT numbers and object density is uniform through the imaged CBCT volume. However, our study provides evidence that this is not the case. Our data systematically demonstrates that the CT numbers are strongly influenced by the anatomical site. We show that the CT number of the same object differed depending on the anatomical location in which it was imaged. Specifically, the CT numbers of phantom objects were higher when placed within the anterior region of the jaw and lower in the posterior regions. This finding underscores an important limitation of using CT numbers on CBCT scans—bone with similar mineralization would yield different CT numbers depending on its anatomical location. Furthermore, our results also show that the magnitude of variation also differed considerably depending on the x-ray attenuation of the object. Objects with low x-ray attenuation suffered greater variability than objects with high attenuation. Overall, our results conclusively demonstrate that CT numbers are not consistent through the image volume, and thus, mathematical equations cannot be easily applied to accurately derive HU through the imaged volume. Perhaps, future development of such algorithms should take into account the discrepancy in CT numbers based on anatomical location.
      There are several factors that could contribute to the anatomical location–based variation of CT numbers on CBCT scans.
      • Nackaerts O.
      • Maes F.
      • Yan H.
      • Couto Souza P.
      • Pauwels R.
      • Jacobs R.
      Analysis of intensity variability in multislice and cone beam computed tomography.
      Most importantly is the concept of “exomass”—in CBCT images, the entire craniofacial skeleton is not included in the image volume.
      • Bryant J.A.
      • Drage N.A.
      • Richmond S.
      Study of the scan uniformity from an i-CAT cone beam computed tomography dental imaging system.
      Thus, there is a significant amount of the patient's tissue that attenuates x-radiation but is not included in the imaged volume. However, typical CBCT reconstruction algorithms assume that the x-ray attenuation takes place only within the imaged volume. A second cause of variation is the amount of scattered radiation.
      • Hunter A.K.
      • McDavid W.D.
      Characterization and correction of cupping effect artefacts in cone beam CT.
      Given the differences in the tissue thickness at the anatomical sites, it is conceivable that the noise from the scattered radiation might contribute, in part, to this inconsistency in the CT numbers. Both the above causes of inconsistency in CT number calculation vary between patients and perhaps also between the scans for each patient. Therefore, it is not practical to apply a mathematical correction to account for these factors. These concepts also highlight a limitation of studies that have used radiographic phantoms—such phantoms typically consist of materials with different densities, encased within a cylinder of homogenous material, usually poly-methyl-methacrylate or water. Frequently, such phantoms are smaller than the FOV and would not suffer the exomass effect encountered in patient imaging.
      • Bryant J.A.
      • Drage N.A.
      • Richmond S.
      Study of the scan uniformity from an i-CAT cone beam computed tomography dental imaging system.
      Our studies were done using the NewTom 3G unit, which uses an image intensifier as the x-ray detector, and the NewTom 5G unit, which uses a flat panel detector. Thus, our results are applicable to all currently available CBCT units. Another factor to consider in interpreting our result is that the kilovolt peak (kVp) used was of 110 for both units. Some CBCT units permit the use of a higher or lower kVp. Notably, lower kVp beams would be more susceptible to artifacts from beam hardening, and plausibly, these imaging protocols may exhibit greater anatomical location–based variation in CT numbers. Similarly, the effect of milliamperage, which differs between different units and between FOVs within the same unit, may also contribute to the inhomogeneity of CT numbers.
      • Katsumata A.
      • Hirukawa A.
      • Noujeim M.
      • Okumura S.
      • Naitoh M.
      • Fujishita M.
      • et al.
      Image artifact in dental cone-beam CT.
      In conclusion, the relationship between CT numbers and object density is not uniform throughout the dental arch. Given this nonuniformity, simple comparison of absolute CT numbers measured at different anatomical locations may be misleading.

      References

        • Tetradis S.
        • White S.C.
        A decade of cone beam computed tomography.
        J Calif Dent Assoc. 2010; 38: 24-26
        • White S.C.
        • Mallya S.M.
        Update on the biological effects of ionizing radiation, relative dose factors and radiation hygiene.
        Aust Dent J. 2012; 57: 2-8
        • Naitoh M.
        • Aimiya H.
        • Hirukawa A.
        • Ariji E.
        Morphometric analysis of mandibular trabecular bone using cone beam computed tomography: an in vitro study.
        Int J Oral Maxillofac Implants. 2010; 25: 1093-1098
        • Isoda K.
        • Ayukawa Y.
        • Tsukiyama Y.
        • Sogo M.
        • Matsushita Y.
        • Koyano K.
        Relationship between the bone density estimated by cone-beam computed tomography and the primary stability of dental implants.
        Clin Oral Implants Res. 2011; 23: 832-836
        • Nackaerts O.
        • Maes F.
        • Yan H.
        • Couto Souza P.
        • Pauwels R.
        • Jacobs R.
        Analysis of intensity variability in multislice and cone beam computed tomography.
        Clin Oral Implants Res. 2011; 22: 873-879
        • Arisan V.
        • Karabuda Z.C.
        • Avsever H.
        • Ozdemir T.
        Conventional Multi-Slice Computed Tomography (CT) and Cone-Beam CT (CBCT) for Computer-Assisted Implant Placement. Part I: Relationship of Radiographic Gray Density and Implant Stability.
        Clin Implant Dent Relat Res. 2012; ([Epub ahead of print; doi:10.1111/j.1708-8208.2011.00436.x])
        • Valiyaparambil J.V.
        • Yamany Y.
        • Ortiz D.
        • et al.
        Bone quality evaluation: comparison of CBCT and subjective surgical assessment.
        Int J Oral Maxillofac Implants. 2012; 27: 1271-1277
        • Mah P.
        • Reeves T.E.
        • McDavid W.D.
        Deriving Hounsfield units using grey levels in cone beam computed tomography.
        Dentomaxillofac Radiol. 2010; 39: 323-335
        • Reeves T.
        • Mah P.
        • McDavid W.
        Deriving Hounsfield units using grey levels in cone beam CT: a clinical application.
        Dentomaxillofac Radiol. 2012; 41: 500-508
        • Sanada S.
        • Kawahara K.
        • Yamamoto T.
        • Takashima T.
        New tissue substitutes representing cortical bone and adipose tissue in quantitative radiology.
        Phys Med Biol. 1999; 44: 107-112
        • Brand J.W.
        • Kuba R.K.
        • Braunreiter T.C.
        An improved head-and-neck phantom for radiation dosimetry.
        Oral Surg Oral Med Oral Pathol. 1989; 67: 338-346
        • Rosset A.
        • Spadola L.
        • Ratib O.
        OsiriX: an open-source software for navigating in multidimensional DICOM images.
        J Digit Imaging. 2004; 17: 205-216
        • Nomura Y.
        • Watanabe H.
        • Honda E.
        • Kurabayashi T.
        Reliability of voxel values from cone-beam computed tomography for dental use in evaluating bone mineral density.
        Clin Oral Implants Res. 2010; 21: 558-562
        • Hohlweg-Majert B.
        • Pautke C.
        • Deppe H.
        • Metzger M.C.
        • Wagner K.
        • Schulze D.
        Qualitative and quantitative evaluation of bony structures based on DICOM dataset.
        J Oral Maxillofac Surg. 2011; 69: 2763-2770
        • Lagravère M.O.
        • Fang Y.
        • Carey J.
        • Toogood R.W.
        • Packota G.V.
        • Major P.W.
        Density conversion factor determined using a cone-beam computed tomography unit NewTom QR-DVT 9000.
        Dentomaxillofac Radiol. 2006; 35: 407-409
        • Bryant J.A.
        • Drage N.A.
        • Richmond S.
        Study of the scan uniformity from an i-CAT cone beam computed tomography dental imaging system.
        Dentomaxillofac Radiol. 2008; 37: 365-374
        • Hunter A.K.
        • McDavid W.D.
        Characterization and correction of cupping effect artefacts in cone beam CT.
        Dentomaxillofac Radiol. 2012; 41: 217-223
        • Katsumata A.
        • Hirukawa A.
        • Noujeim M.
        • Okumura S.
        • Naitoh M.
        • Fujishita M.
        • et al.
        Image artifact in dental cone-beam CT.
        Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2006; 101: 652-657